Sysmex Xn-2000 User Manual Pdf Jun 2026
Used for micro-tainers, urgent STAT samples, or tubes with unreadable barcodes. Select manual mode on the IPU, mix the sample thoroughly, place the tube under the aspiration probe (or closed-tube piercer), and press the start switch. Reagent Management
Approximately 37.8" (W) x 36" (D) x 33.7" (H), weighing roughly 315 lbs. Core Technologies and Principles
Adhering to the maintenance schedule outlined in the user manual protects the instrument warranty and ensures mechanical longevity. Daily Maintenance
The Sysmex XN-2000 utilizes advanced fluorescence flow cytometry, hydrodynamic focusing, and semiconductor laser technology. Understanding the core parameters and components is essential before operating the system. Core Analytical Channels
When you finally open your , here are the five sections you should bookmark immediately: Sysmex Xn-2000 User Manual Pdf
✅
The XN-2000 is a featuring two integrated analytical modules. This configuration allows for a massive throughput of up to 200 samples per hour and provides an automatic backup system; if one module is down for maintenance, the other continues to process samples. Key Detection Methods:
The manual provides instructions for performing daily QC, managing Sysmex SNCS (Sysmex Network Communications System) data, and troubleshooting QC failures to ensure accurate patient results. 5. Maintenance Procedures
Select the manual analysis icon on the IPU screen. Mix the sample manually, open the tube cap, place the tube into the manual aspiration probe micro-adapter, and press the physical aspiration switch. Reagent Management Used for micro-tainers, urgent STAT samples, or tubes
Measures RBC (Red Blood Cells) and PLT (Platelets) using hydrodynamic focusing and electronic resistance. Hb (Hemoglobin) is measured via a SLS-hemoglobin method.
If multiple parameters are out across all levels, open a fresh vial of QC. 6. Flagging and Data Review
If the analyzer flags a clogging error, the manual instructs users to run a special "Shut Down" or "Clean" cycle, followed by checking the sample probe.
Verify that the background check values fall within permissible limits: RBC: HGB: ≤0.1is less than or equal to 0.1 PLT: Step 2: Running Quality Control (QC) Core Technologies and Principles Adhering to the maintenance
Indicates potential presence of malignant cells. Review via a manual peripheral blood smear.
Remove XN-CHECK vials from the refrigerator and allow them to reach room temperature (15–30°C) for at least 15 minutes. Mix gently by inversion according to package inserts.
The XN-2000 is engineered for speed and reliability, featuring a shared sampler and an integrated backup concept that ensures continuous operation even if one module requires maintenance.
Run QC daily, after changing reagent lots, or following major maintenance.